Management of Electrocautery in Patients with Deep Brain Stimulators

Preoperative Preparation

  • The American Society of Anesthesiologists recommends turning off the deep brain stimulator before surgery, using bipolar electrocautery whenever possible, and if monopolar cautery is necessary, positioning the grounding pad to avoid current pathways through the DBS system while using short, intermittent bursts at minimum power settings 1
  • Interrogate the device preoperatively to document baseline settings and ensure proper function, and have the device programmer or manufacturer contact information immediately available 1
  • Ensure temporary external monitoring equipment is available throughout the procedure, and have cardiac monitoring in place from the time of device reprogramming until procedure completion 1

Intraoperative Electrocautery Technique

  • The American Society of Anesthesiologists strongly prefers bipolar electrocautery systems over monopolar, as they confine current flow between the two tips of the forceps and produce minimal electromagnetic interference 1, 2
  • Position the electrosurgical grounding plate so the current pathway does NOT pass through or near the DBS pulse generator or leads, and for head and neck procedures, place the grounding pad on the superior posterior aspect of the shoulder contralateral to the generator position rather than the standard thigh placement 1
  • Use the absolute minimum power settings necessary for hemostasis, and apply only short, intermittent, and irregular bursts rather than continuous activation 1, 3

Postoperative Management

  • Re-interrogate the device to confirm proper function and verify settings have not changed, and document any changes in device parameters 1
  • Assess for signs of electrode heating injury, which may manifest as elevated pacing or sensing thresholds in the days following surgery, and monitor for device malfunction including reprogramming, inhibition, or noise reversion mode 2

Critical Pitfalls to Avoid

  • Never position grounding pads in standard locations without considering the DBS current pathway, as this is the most common cause of electromagnetic interference complications 1
  • Never use continuous or high-power cautery settings when any electrosurgical device is necessary near a DBS system 1, 3
  • Never proceed without having device programming capability immediately available, as emergency reprogramming may be necessary if malfunction occurs 1

Special Considerations

  • The risks are similar to those seen with cardiac rhythm management devices, including device reprogramming, inhibition, local electrode heating causing brain tissue damage, and noise reversion mode 2
  • Consider postponing elective procedures until device management can be optimized, and optimize grounding pad positioning even more carefully if the patient is highly dependent on continuous DBS therapy and turning it off is not feasible 1