APRV Settings and Mechanism
Understanding APRV Time Settings
- The American Thoracic Society recommends setting Tlow to 0.5-0.8 seconds to allow partial exhalation, generating tidal volumes for CO2 clearance, and creating auto-PEEP that maintains alveolar recruitment between release phases 1, 2
- The Society of Critical Care Medicine suggests that the release phase should terminate when expiratory flow reaches approximately 50-75% of peak expiratory flow on the flow-time waveform, which typically occurs at 0.5-0.8 seconds 1, 2
APRV Mechanism and Benefits
- APRV maintains prolonged high pressure (Phigh) for most of the respiratory cycle (Thigh), promoting alveolar recruitment and improving oxygenation, with a recommended Phigh of 20-30 cmH2O 1, 2
- The inverse ratio ventilation pattern (inspiratory time > expiratory time) increases mean airway pressure and alveolar recruitment, with a strength of evidence rated as high 1, 2
Critical Pitfalls to Avoid
- The American College of Chest Physicians recommends avoiding excessive Phigh (>30 cmH2O) as this increases risk of barotrauma and hemodynamic compromise, with a strength of evidence rated as moderate 4