Anesthesia System Configuration and Safety
Configuring Anesthesia Delivery for Specific Clinical Scenarios
- The American Society of Anesthesiologists recommends using an open gas delivery device, such as a face mask or nasal cannula, only if the patient is not oxygen-dependent and does not require moderate or deep sedation, when managing fire risk during head or neck surgery with an ignition source 1
- Before activating any ignition source around the face, head, or neck, stop supplemental oxygen delivery or reduce to the minimum required to avoid hypoxia, wait 1-3 minutes between decreasing oxygen flow and approving ignition source activation, and if using nitrous oxide, stop it and wait 1-3 minutes before ignition, as recommended by the American Society of Anesthesiologists 1
- A sealed gas delivery device, such as a cuffed tracheal tube or laryngeal mask, should be used instead of an open device if moderate or deep sedation is required or if the patient exhibits oxygen dependence, according to the American Society of Anesthesiologists 1
Preparing Anesthesia Machines for Trigger-Free Anesthesia
- The British Journal of Anaesthesia recommends removing vaporizers before flushing the machine, replacing breathing circuits, fresh gas hose, and soda lime, and flushing with maximum fresh gas flow at ≥10 L/min for the manufacturer-recommended duration to prepare a machine for trigger-free anesthesia 2, 3
- During flushing, set tidal volume at 600 mL and frequency at 15 bpm if using mechanical ventilation, as recommended by the British Journal of Anaesthesia 2, 3
- Consider using activated charcoal filters if available to reduce volatile agent concentrations to <5 ppm in 2-3 minutes, according to the British Journal of Anaesthesia 2
Safety Considerations
- The American Society of Anesthesiologists advises against routine supplemental oxygen delivery via open devices, such as nasal cannula or face mask, in the operating room, as this creates unnecessary fire risk 1