Guidelines for Proper Oxygen Device Delivery
Oxygen Delivery Principles
- The proper oxygen device should be selected based on the patient's clinical condition, target oxygen saturation range, and risk of hypercapnic respiratory failure, with nasal cannulae being the first-choice delivery device for most patients requiring home or hospital oxygen therapy 1, 2
- Oxygen is a treatment for hypoxemia, not breathlessness, and should be administered by staff trained in oxygen administration 3
- Oxygen should be prescribed according to a target saturation range, and those administering oxygen therapy should monitor the patient to keep within this target range 3
- For most acutely ill patients without risk of hypercapnic respiratory failure, aim for a target saturation of 94-98% 4
- For patients with COPD or other risk factors for hypercapnic respiratory failure, aim for a target saturation of 88-92% 5, 4
Oxygen Delivery Devices
- Nasal cannulae are recommended as the first-choice delivery device for most patients requiring oxygen therapy 1, 2
- The flow rate for nasal cannulae should be adjusted between 2-6 L/min for medium-concentration oxygen therapy to achieve desired target saturation 2
- For patients at risk of hypercapnic respiratory failure, use 1-2 L/min with nasal cannulae 2, 6
- Simple face masks deliver oxygen concentrations of 40-60% at flow rates of 5-10 L/min, but flow rates should never be below 5 L/min due to risk of carbon dioxide rebreathing 2, 7
- Venturi masks are recommended for patients requiring precise control of FiO₂ and can deliver constant FiO₂ of 24%, 28%, 31%, 35%, 40%, and 60% oxygen 2, 7
- High-concentration reservoir masks deliver oxygen concentrations of 60-90% at a flow rate of 15 L/min, but are not suitable for patients at risk of hypercapnic respiratory failure 2, 6, 7
Monitoring Oxygen Therapy
- Oxygen saturation should be observed for at least 5 minutes after starting oxygen therapy or after changing concentration 1
- Record oxygen saturation, delivery system, and flow rate on the patient's monitoring chart 1, 3
- Most stable patients should have saturation measured after 1 hour of oxygen therapy and then four-hourly 1
- Patients with signs of critical illness should have oxygen saturation monitored continuously 8
- Pulse oximetry should be available in all locations where emergency oxygen is used 3
- For patients at risk of hypercapnic respiratory failure, arterial or arterialised capillary blood gases should be measured 30-60 minutes after an increase in oxygen therapy 8
Special Considerations
- For perioperative and postoperative care, medium-concentration masks or nasal cannulae are usually sufficient, except for COPD patients 6
- In emergency departments, consider the risk of hypercapnic respiratory failure when selecting devices 6
- For general and respiratory wards, device selection depends on expected duration of treatment, type of respiratory illness, breathing pattern, need for humidification, risk of CO₂ retention, and patient compliance 6
Practical Implementation
- Ensure all clinicians prescribing oxygen have appropriate training 8
- Every hospital should have a training program to ensure clinical staff are familiar with oxygen administration policies 8
- For home oxygen therapy, ensure patients/caregivers understand safety recommendations including proper storage of cylinders, no smoking near oxygen, and avoiding naked flames 6