Electromagnetic Interference with Pacemakers during Bipolar Cautery

Risk Reduction and Management

  • The European Society of Cardiology recommends using bipolar cautery systems, which are less hazardous for patients with pacemakers, as they reduce the risk of electromagnetic interference (EMI) compared to unipolar systems 1, 3
  • Bipolar cautery confines the electrical current between the two electrodes of the cautery device, significantly reducing the amount of current that can potentially interfere with pacemaker function, as suggested by the American College of Cardiology 2, 4
  • The American College of Cardiology notes that the probability of adverse interactions has fallen dramatically due to improved pacemaker design and the almost universal use of bipolar leads, but EMI can still occur 2, 5

Factors Affecting EMI Risk

  • The American College of Cardiology states that the likelihood and potential clinical impact of EMI with pacemakers depends on several factors, including whether the patient is pacemaker-dependent 2, 4
  • The European Heart Rhythm Association suggests that bipolar leads provide better protection against EMI than unipolar leads 2, 6
  • The American College of Cardiology recommends considering the relative distance from and orientation of the cautery relative to the pacemaker and leads, as well as the surgical site, when assessing EMI risk 2, 6

Recommendations for Perioperative Management

  • The European Society of Cardiology recommends pre-operative reprogramming of the pacemaker to an asynchronous or triggered mode for pacemaker-dependent patients, and ensuring provisions for immediate activation of asynchronous pacing for non-pacemaker-dependent patients 1, 3
  • The American College of Cardiology suggests keeping cautery use and power output to the minimum required, applying cautery in short bursts, and not in close proximity to the device 1, 3
  • The European Society of Cardiology recommends monitoring ECG and pulse oximetry during the procedure, and interrogating the device after the procedure if cautery was used near the pacemaker or leads 1, 3

Potential EMI Effects on Pacemakers

  • The American College of Cardiology notes that even with bipolar cautery, adverse effects such as temporary or permanent resetting to a backup pacing mode, temporary or permanent inhibition of pacemaker output, increased pacing rate, ICD firing, and myocardial injury at the lead tip can still occur 2, 4