Safety Considerations for Non-Interlocked Vaporizers

Key Safety Risks

  • Non-interlocked vaporizers allow multiple vaporizers to be turned on simultaneously, which can result in uncontrolled and potentially lethal concentrations of anesthetic vapors being delivered to patients 2

Essential Safety Precautions

  • Ensure that vaporizers are correctly fitted to the anesthetic machine with any locking mechanism fully engaged 1, 3
  • Verify that vaporizer control knobs rotate fully through their entire range 1
  • Check that vaporizers are adequately filled but not overfilled, with filling ports tightly closed 1
  • Perform manual leak testing on basic machines by setting oxygen flow at 5 L/min with vaporizer turned off, temporarily occluding the common gas outlet, and confirming no leaks from any part of the vaporizer 1, 2, 4

Critical Pitfalls to Avoid

  • Never operate multiple vaporizers simultaneously as this can lead to unpredictable and potentially dangerous anesthetic concentrations 2

Modern Safety Solutions

  • Modern anesthetic workstations often include automatic vaporizer interlocking systems and vaporizer integrity testing 2, 4