Mechanical Ventilation Maneuvers

Assessment of Respiratory Mechanics

  • The American Thoracic Society recommends using end inspiratory holds to measure plateau pressure and assess respiratory system compliance, with a strength of evidence rated as high 1, 3
  • The American College of Chest Physicians suggests that driving pressure, calculated as plateau pressure minus PEEP, represents the pressure needed to deliver the tidal volume, with a moderate strength of evidence 1, 3
  • Respiratory system compliance is derived from tidal volume divided by driving pressure, with a high strength of evidence 1, 3

Clinical Applications of Inspiratory Holds

  • The American College of Cardiology recommends assessing risk of ventilator-induced lung injury by keeping plateau pressures below 30 cmH₂O in ARDS patients and below 25 cmH₂O in non-ARDS patients, with a high strength of evidence 4
  • The Society of Critical Care Medicine suggests guiding PEEP titration using the overdistension-collapse method during inspiratory holds at different PEEP levels, with a moderate strength of evidence 3
  • Distinguishing between airway resistance and lung compliance problems can be achieved by measuring the difference between peak pressure and plateau pressure, with a low strength of evidence 1

Technical Requirements for Inspiratory Holds

  • The American Association for Respiratory Care recommends performing end inspiratory holds during volume-controlled ventilation or pressure-controlled ventilation with adequate equilibration time (>0.5 seconds), with a high strength of evidence 3
  • Patients must be passive (not actively breathing) to avoid artifacts from respiratory muscle activity, with a high strength of evidence 1, 2

Assessment of Auto-PEEP

  • The American Thoracic Society recommends using end expiratory holds to detect intrinsic PEEP (auto-PEEP), which represents trapped gas and incomplete lung emptying, with a high strength of evidence 2, 6
  • Auto-PEEP can reach 10-15 cmH₂O in severe obstructive disease, requiring enormous patient effort to initiate breaths, with a moderate strength of evidence 6
  • Auto-PEEP causes hemodynamic instability by decreasing venous return and cardiac output through increased intrathoracic pressure, with a high strength of evidence 2, 6

Clinical Significance of Auto-PEEP

  • Breath stacking from auto-PEEP leads to barotrauma, tension pneumothorax, and hypotension, with a high strength of evidence 7, 8
  • The American College of Chest Physicians suggests guiding application of external PEEP to counterbalance auto-PEEP (typically 50-85% of measured auto-PEEP, never exceeding it), with a moderate strength of evidence 6, 4

Technical Requirements for Expiratory Holds

  • The Society of Critical Care Medicine recommends using the end-expiratory occlusion technique as the gold standard for measuring auto-PEEP, with a high strength of evidence 2, 4
  • Patients must be completely passive during measurement; expiratory muscle recruitment invalidates the measurement, with a high strength of evidence 2, 4