Immunohistochemical Markers for Confirming Plasma Cell Clonality in Multiple Myeloma

Essential Markers for Plasma Cell Identification

  • CD38, CD138, and CD45 should all be included in at least one tube for plasma cell identification and enumeration 1, 2
  • The primary gating for plasma cell identification should be based on CD38 vs. CD138 expression 2
  • These markers allow for the unequivocal identification of plasma cells among other hematopoietic cells 3

Core Markers for Clonality Assessment

  • Cytoplasmic kappa and lambda light chain expression is critical to demonstrate clonality of plasma cells 4, 5
  • Assessment of cytoplasmic κ/λ expression is important at presentation and appropriate for assessment of stringent complete remission according to IMWG criteria 5

Minimal Panel for Detecting Abnormal Plasma Cells

  • CD19 and CD56 form the minimal panel for detection of abnormal plasma cells 2, 3
  • CD19 negativity is a hallmark of neoplastic plasma cells (normal plasma cells are typically CD19+) 4

Preferred Extended Panel

  • CD20, CD117, CD28, and CD27 should be included in a preferred panel 2, 3, 4
  • CD117 (c-kit) expression is common in myeloma cells but absent in normal plasma cells 4
  • CD20 is typically negative in plasma cells but may be expressed in a subset of myeloma cases 4

Technical Considerations

  • For light chain assessment, bone marrow samples should be washed twice in buffered saline solution prior to assessment of cytoplasmic immunoglobulin expression to remove cytophilic immunoglobulin 5
  • Standard commercial fixation and permeabilization kits are suitable for assessment of cytoplasmic kappa/lambda detection 5
  • At least 100 neoplastic plasma cell events should be acquired for accurate enumeration when using flow cytometry to assess light chain expression 6

Sample Quality Assessment

  • The sample is suitable for quantitative analysis if normal plasma cells (CD19+CD56- and/or polyclonal) are detectable 4
  • If normal plasma cells are not detected, sample quality should be assessed by morphological evaluation or additional flow cytometry for normal marrow elements 4
  • Discrepancies between the percentage of plasma cells detected by flow cytometry and morphology are primarily related to sample quality 1

Pitfalls and Caveats

  • The demonstration of phenotypically abnormal plasma cells is more sensitive and specific for detecting residual disease than clonality assessment alone 5
  • No single marker can systematically differentiate neoplastic plasma cells from their normal counterparts, necessitating a panel approach 4