MRI Protocol for Quantifying Iron Concentration in Liver and Heart

Liver Iron Quantification Protocol

  • The American College of Radiology recommends R2*-based relaxometry as the gold standard first-line method for liver iron concentration quantification with high-level evidence supporting its use 1, 3
  • For liver iron quantification, both 1.5T and 3T field strengths are acceptable, with 1.5T preferred for suspected severe iron overload due to better dynamic range 2, 4, 5
  • The maximum reliable liver iron concentration (LIC) is ~40 mg/g at 1.5T and ~26 mg/g at 3T 2, 4
  • Protocol parameters for R2*-based relaxometry include short echo times (<1 msec), short echo spacing (<1 msec), 6-12 echo times, and noise correction 1, 3, 6
  • Fat correction is required to avoid fat- and noise-related biases 1, 3, 6

Alternative Methods for Liver Iron Quantification

  • R2-based relaxometry (FerriScan) is a second-line option with regulatory approval, requiring 1.5T field strength and longer acquisition time (10-20 minutes) 7, 8
  • Signal Intensity Ratio (SIR) is a third-line alternative, well validated at 1.5T, but limited to LIC <19.5 mg/g and without regulatory approval 7, 8, 9

Cardiac Iron Quantification Protocol

  • No cited facts are available for cardiac iron quantification protocol

Combined Liver and Heart Protocol Order

  • The recommended order details for MRI liver and cardiac iron quantification with R2* relaxometry include specifying sequence type, field strength, echo parameters, and corrections 1, 2, 3, 4, 6

Critical Pitfalls to Avoid

  • Field strength matters for severe overload, as 3T may fail to quantify LIC >26 mg/g due to dynamic range limitations 2, 4, 5
  • Ensure proper corrections, as uncorrected methods for fat and noise introduce significant bias 1, 3, 6

Reporting Requirements

  • The radiology report must include the specific MRI method used, literature citation describing the calibration method, laboratory reference values used for LIC calibration, and quantitative values in both mg/g and μmol/g 1, 3, 10