Blood Glucose Meter Accuracy Standards

Primary Considerations for Meter Selection

  • The Diabetes Technology Society recommends using only meters that meet FDA accuracy standards, which require 95% of readings within 15% for home use and 99% within 20% across the usable glucose range 2, 3
  • The American Diabetes Association suggests that meters should be selected based on their ability to meet current FDA accuracy standards, with ongoing verification through independent testing 1, 2

Critical Accuracy Standards

  • The FDA maintains stricter standards than ISO for hospital use, requiring 95% of readings within 12% for glucose ≥75 mg/dL and 95% within 12 mg/dL for glucose <75 mg/dL 5, 3
  • The International Organization for Standardization (ISO) standard for hospital use requires 95% of readings within 15% for glucose ≥100 mg/dL and 95% within 15 mg/dL for glucose <100 mg/dL 5

Practical Recommendations for Meter Use

  • The European Association for the Study of Diabetes recommends ensuring the meter is appropriate for the clinical setting, with professional-use meters for hospitals and home-use meters for outpatient management 3
  • The American College of Clinical Endocrinologists suggests matching meter technology to patient factors, such as using glucose dehydrogenase-based meters for patients with abnormal oxygen saturation 3
  • The Centers for Disease Control and Prevention (CDC) advises verifying test strips are unexpired and properly stored, purchased only from pharmacies or licensed distributors 4, 2