Target Controlled Infusion of Propofol
Basic Principles
- TCI systems use an infusion pump connected to a computer programmed with drug-specific, population-based pharmacokinetic models to deliver propofol 1.
- The computer calculates the necessary infusion rate to achieve the target drug concentration and signals the pump to deliver the appropriate amount 2.
- Unlike constant-rate infusions which may lead to drug accumulation, TCI uses mathematical models to calculate the initial dose needed and then makes appropriate adjustments to maintain the desired level 1.
Types of TCI Systems
- In open-loop systems, the physician selects a target concentration, and the system delivers propofol to achieve that concentration 3.
- The pharmacokinetic model adjusts for patient characteristics including age, sex, body weight, and comorbid disease 3.
- Continuous monitoring is required as individual responses to a given drug concentration will differ due to pharmacodynamic variability 3.
- Closed-loop systems use real-time feedback from measures of drug effect to regulate the delivered drug concentration 3.
- The operator selects the desired level of drug effect, and the computer regulates the infusion to maintain that set-point 3.
- These systems provide more individualized sedation, potentially reducing undersedation and oversedation 3.
Clinical Application of TCI Propofol
- For sedation during procedures, initial plasma concentrations typically range from 0.8-4 μg/mL depending on the desired level of sedation 4, 5.
- Respiratory depression, hypoxia, and apnea can occur with TCI propofol, requiring continuous monitoring 4, 5.
- Computer-assisted personalized sedation systems can enhance safety by automatically reducing or stopping infusions in response to clinical or physiological indications of oversedation 6.
Target Concentration for Dental Sedation
Monitoring and Sedation Management
- Continuous pulse oximetry, heart rate, blood pressure, and clinical responsiveness assessment are mandatory throughout sedation, and maintain patient responsiveness to verbal commands or gentle tactile stimulation to confirm moderate sedation, as recommended by the American Gastroenterological Association 7
- The American Society of Anesthesiologists suggests that essential parameters such as pulse oximetry, heart rate, and blood pressure should be continuously monitored during sedation, with a target concentration of 1.0 μg/mL propofol effect-site concentration 7