Ultrasound in Ophthalmology

Introduction to Ultrasound Biomicroscopy

  • Ultrasound biomicroscopy utilizes high-frequency sound waves (35-80 MHz range) to generate real-time cross-sectional images of the anterior segment, allowing for continuous real-time imaging with high spatial resolution and sensitivity, as noted by the American Academy of Ophthalmology 1

Advantages and Capabilities

  • The DxU 850m Iris provides superior visualization of anterior segment structures even through opaque corneas, and can image virtually all anterior segment anatomy, including cornea, iridocorneal angle, anterior chamber, iris, ciliary body, and lens, according to the American Academy of Ophthalmology 2
  • The primary advantage of the DxU 850m Iris over light-based imaging devices is its ability to penetrate through corneal opacities, anterior segment, or vitreous opacities, making it invaluable for evaluating patients with corneal edema or opacification, as recommended by the American Academy of Ophthalmology 2
  • High-frequency ultrasound (35-80 MHz) provides significantly improved resolution compared to conventional 10 MHz ultrasound, as reported by the American Academy of Ophthalmology 2

Clinical Applications

  • Ultrasound biomicroscopy is particularly useful in identifying root causes of corneal edema or opacity, and can detect Descemet membrane ruptures or detachments, central stromal clefts in keratoconic hydrops or trauma, and retrocorneal membranes hidden by corneal edema or scarring, according to the American Academy of Ophthalmology 3
  • The system aids in surgical planning for corneal procedures by providing detailed structural information, and is especially helpful in congenital and traumatic cases, as noted by the American Academy of Ophthalmology 2
  • Ultrasound biomicroscopy can guide endothelial keratoplasty management in the immediate postoperative period, and can locate small anterior segment foreign bodies that are difficult to detect by slit-lamp examination or anterior segment OCT, as reported by the American Academy of Ophthalmology 3

Imaging Techniques and Considerations

  • Examination with ultrasound biomicroscopy begins with conventional B-mode imaging to identify target structures, and the transducer is then held motionless while switching to contrast-specific imaging mode, as described by the American Academy of Ophthalmology 1
  • Proper probe positioning at a 90° angle is essential for consistent measurements, as recommended by the American Academy of Ophthalmology 2
  • The American Academy of Ophthalmology recommends using frequencies of 35-80 MHz for anterior eye segment imaging, which provides the best detail 2
  • High frequency ultrasound is used for anterior segment imaging with high detail, providing superior resolution for superficial structures, particularly valuable for imaging the cornea, anterior chamber angle, iris, ciliary body, and lens 4, 2
  • When selecting the appropriate HFUS frequency, consider the depth of the target structure, balance the need for resolution against penetration requirements, and choose frequencies based on the specific clinical application, such as 20 MHz and above for dermal applications, 35-80 MHz for anterior eye segment imaging, and 10 MHz and above for joint imaging 4, 2, 5
  • The American College of Emergency Physicians recommends documenting the indication for the procedure, recording the views obtained, describing the structures studied, providing interpretation of findings, and noting any limitations or impediments to performing a complete exam 6

Ultrasound Modalities

  • A-scan ultrasound uses 7-11 MHz transducers, with 7-8 MHz for determining echogenicity of ocular tumors and 10-11 MHz for measuring axial length of the eye, according to the American Institute of Ultrasound in Medicine 2
  • B-scan ultrasound operates at 10-20 MHz for measuring posterior ocular tumors, as recommended by the American Academy of Ophthalmology 2
  • Ultrasound biomicroscopy (UBM) operates at 40-100 MHz to evaluate anterior ocular structures, according to the American Academy of Ophthalmology 2

Safety and Limitations

  • Plane-wave modalities are well within FDA safety guidelines for ocular use, according to the Food and Drug Administration 2
  • Using too high a frequency for posterior segment evaluation can result in inadequate penetration, while using too low a frequency for anterior segment structures can result in insufficient resolution, as warned by the American Academy of Ophthalmology 2
  • Improper probe positioning can affect measurement accuracy, and failing to adjust frequency based on the specific ocular structure being examined can lead to suboptimal image quality, according to the American College of Emergency Physicians 6